Differential Calorimeter for Cement and Concrete - ToniCAL

Differential Calorimeter for Direct Determination of the Heat of Hydration (J/g) of Cement

ToniCAL - Differential Calorimeter for Direct Determination of the Heat of Hydration (J/g) of Cement

ToniCAL differential calorimeter is an instrument used to analyze the setting behaviour of building materials such as cement and mortar. It is able to determine the rate of heat generation with respect to time with the rigid control of a computer control system.

ToniCAL is made up of two main elements, the calorimeter and the control device.

Calorimeter

The calorimeter is designed with a steel housing to protect the samples from environmental heat fluctuation as well as specimen-to-specimen heat transfer. Each measuring cell is also fitted with a heating element and thermoelectric conductive measuring chains for heat output readings.


Control Device

From the beginning of the test, the control device operates the procedure with a direct interface. The control stabilizes the thermal signal and amplifies it to be a measurable voltage. It also monitors the internal measuring cell temperature. All data is logged on an internal drive for later download.


Special Features:
  • Fully-automatic test procedure
  • Evaluation and documentation of test results via computer
  • Rigid, durable measuring instrument for industrial and laboratory use
  • Simple operation
  • Self-calibrating

Technical Data

Voltage

230

V

Variance

±10

%

Power Frequency

50/60

Hz

Power Consumption

<250

VA

Dimensions

control unit

325 x 310 x 280

mm

calorimeter

600 x 375

mm

Weight

control unit

25

kg

calorimeter

36.5

kg

Integrated Measuring Cell

Warm-up Time by Air Control

<8

h

Ambient Temperature

20

ºC

Ambient Temperature Area

+3 … 50

ºC

Working Temperature Area

Ambient temp +5K

ºC

Initial Weight:

solid

10

G

liquid

5

G

Detection Barrier

0.2

J/g


ToniCAL Trio

Triple Chamber Computer Controlled Differential Scanning Calorimeter for Direct Determination of the Heat of Hydration (J/g) of Cement

ToniCAL Trio - Triple Chamber Computer Controlled Differential Scanning Calorimeter for Direct Determination of the Heat of Hydration (J/g) of Cement

ToniCAL TRIO is an extension of ToniCAL that contains three independent chambers for calorimetry allowing a greater productivity in test results. The extra chambers allows for parallel independent time varying use of each test specimen. This allows for more accurate calorimetry measurements including determination of background noise.

Special Features:
  • Rigid durable apparatus for industrial and laboratory use
  • High resolution and sensitivity
  • Simple handling
  • Test procedure with extremely high information for product development
  • High reproducibility
  • Recording and interpretation of the test results in numeric and graphic recording and documentation
  • Self calibration system