TR Tester ET 01
For determination of Low Temprature Retraction

TR Tester, ET 01, for determination of low temperature characteristics
by the temperature retraction procedure according to ISO 2921
and ASTM D-1329.
TR Tester, has 6 test stations,
is computerized and performs the test automatically after the
cooling media has been cooled down and the samples has been
mounted. An automatic release of the samples, after the pre
cooling period is included.
The computer controls both the temperature
rise and measures the length change of the samples. The results
are displayed in graph and TR10, TR30, TR50 and TR70 values
are calculated. The result can also be presented as a table
with length change versus temperature. The TR-values and the
table values can be exported to other software such as spreadsheets.
As an extra option, an automatic cooling system with liquid
Nitrogen can be supplied. |

 |
Technical
Specifications |
Temperature Control |
| Temperature Range |
-80°C
to +30°C |
|
Accuracy |
±0.5°C |
| Resolution |
±0.1°C |
| Sensor |
PT 100, 1/3 DIN |
Length Measurement |
| Indicators |
6 pc digital encoders
|
| Resolution |
0.04mm |
| Measuring Range |
0mm - 150mm
|
Test Rig |
| Rig Material |
Stainless steel and aluminium |
| Specimen Sizes |
25mm to 100mm
|
| Max Extension |
250% (with 50mm test piece) |
| Test Stations |
6
|
Other Specifications |
| Dimensions, External |
575mm x 1100mm x 400mm (w x h x d) 575mm x 1420mm x 400mm (w x h x d) Raised Test Rig |
| Weight |
Approx 40Kg
|
| Cooling Liquid Volume |
About 8l |
| Voltage |
220-240 VAC 50/60 Hz (alt 110-120 VAC)
|
| Power |
600W |
| Air Supply |
4 - 5 Bar
|
| Computer Connection |
COM-port (Serial, RS 232C) |
 |
- The
casing consists of steel, painted with epoxy paint.
- The
rig is raised with the help of a pneumatic cylinder.
- Temperature
controller with 0,1 C setpoint and RS232 interface.
- Solid
state rely for safe control.
- Temperature
sensor close to the samples.
- Run-time
meter.
- ET
021 Windows software for the Gehman Tester is included A
Pentium 133 with 64 MB RAM or better is required.
Options